CARESTREAM MEDICAL FDA Inspection Record
CARESTREAM MEDICAL is an FDA-registered facility (FEI 3010396502) located in Sanford, FL, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 6, 2026: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest May 15, 2018
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 6, 2026 | Voluntary Action Indicated (VAI) | Label to bear a unique device identifier (21 CFR 801.20(a)) QMS not documented and maintained adequately (ISO 13485:2016 Clause 4.1.1) Top management did not ensure requirements are determined and met (ISO 13485:2016 Clause 5.2) Design and development procedures not documented (ISO 13485:2016 Clause 7.3.1) and 3 more |
| Apr 27, 2022 | No Action Indicated (NAI) | Design control - no procedures (21 CFR 820.30(a)) Design history file (21 CFR 820.30(j)) Complaints (21 CFR 820.198(a)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 15, 2018 | Class II | Z-2887-2018 | CAREstream ProNox Nitrous Oxide Delivery System labeled as the following:
a. CMI-0100… Two analgesic gas delivery system oxygen/nitrous oxide mixers were reported to have delivered higher levels of nitrous oxide than… |
FDA record score
7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Feb 6, 2026 | 6 x 1 | 6 |
| Class II recall Z-2887-2018, this FEI as the recalling firm | May 15, 2018 | 4 x 0.25 | 1 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing CARESTREAM MEDICAL as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check CARESTREAM MEDICAL yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- CAREstream Medical Ltd
- Location
- Sanford, FL (US)
- Source
- device