Coapt LLC FDA Inspection Record
Coapt LLC is an FDA-registered facility (FEI 3010605876) located in Chicago, IL, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Oct 28, 2024: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (2 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jan 8, 2016. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Oct 2, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 28, 2024 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Documentation (21 CFR 820.100(b)) |
| Nov 2, 2022 | Voluntary Action Indicated (VAI) | Document change records, maintained. (21 CFR 820.40(b)) Justification for not reporting (21 CFR 806.20(b)(4)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Dec 11, 2019 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Sep 20, 2016 | No Action Indicated (NAI) | No citations listed |
| Nov 14, 2014 | No Action Indicated (NAI) | Quality policy and objectives (21 CFR 820.20(a)) Quality plan (21 CFR 820.20(d)) Quality System Procedures (21 CFR 820.20(e)) Documentation (21 CFR 820.80(e)) and 4 more |
Warning letters
- Jan 8, 2016: Warning letter listed on FDA's Data Dashboard. View compliance record (448309).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 2, 2023 | Class II | Z-0203-2024 | Coapt Dome Electrode, Gen2 system, cutaneous electrode Patient may experience minor burn or blistering if device is exposed to direct sunlight and/or warm temperatures for over 12 hours or… |
| Oct 2, 2023 | Class II | Z-0204-2024 | Coapt ControlSeal Electrode, cutaneous EMG Electrode Patient may experience minor burn or blistering if device is exposed to direct sunlight and/or warm temperatures for over 12 hours or… |
FDA record score
12 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Oct 28, 2024 | 6 x 0.6 | 3.6 |
| Inspection classified VAI | Nov 2, 2022 | 6 x 0.6 | 3.6 |
| Class II recall Z-0203-2024, this FEI as the recalling firm | Oct 2, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0204-2024, this FEI as the recalling firm | Oct 2, 2023 | 4 x 0.6 | 2.4 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jan 8, 2016 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Coapt LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.