INNOVATIVE STERILIZATION TECHN FDA Inspection Record

FEI 3010772758 · Inspections, Form 483s, warning letters and import refusals

INNOVATIVE STERILIZATION TECHN is an FDA-registered facility (FEI 3010772758) located in Dayton, OH, US. RegDataLab tracks 6 FDA inspections and 2 compliance actions at this facility.

What FDA found at this site

Latest inspection
Aug 1, 2024: No Action Indicated (NAI)
Inspections on record
6 (2 OAI, 3 VAI, 1 NAI)
Last OAI inspection
Nov 4, 2015
Warning letters
2 listed by FDA for this FEI, latest Mar 20, 2019
Recalls as the recalling firm
1, latest Feb 3, 2014

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 1, 2024No Action Indicated (NAI)No citations listed
Jan 13, 2022Voluntary Action Indicated (VAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Apr 12, 2017Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
Lack of System for Determining MDR Events (21 CFR 803.17(a)(2))
Nov 4, 2015Official Action Indicated (OAI)Procedures not adequately established or maintained (21 CFR 820.40)
Design control - no procedures (21 CFR 820.30(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b))
Aug 13, 2015Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 1 more
Dec 22, 2014Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Procedures not adequately established or maintained (21 CFR 820.40)
Lack of Written MDR Procedures (21 CFR 803.17)

Warning letters

Letters whose text states this FEI. Many letters do not state one, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 3, 2014Class IIZ-0153-2016OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container…
The wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at 270 degrees F instead of Steam Gravity 17…

FDA record score

13 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

8 records behind this number
RecordDatePoints x age weightAdds
Warning letter, no close-out letter listedMar 20, 201930 x 0.257.5
Inspection classified VAIJan 13, 20226 x 0.63.6
Inspection classified VAIApr 12, 20176 x 0.251.5
Warning letter on FDA's dashboard, which does not show close-outsMar 2, 201610 years or older, so it adds no points0
Inspection classified OAINov 4, 201510 years or older, so it adds no points0
Inspection classified OAIAug 13, 201510 years or older, so it adds no points0
Inspection classified VAIDec 22, 201410 years or older, so it adds no points0
Class II recall Z-0153-2016, this FEI as the recalling firmFeb 3, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing INNOVATIVE STERILIZATION TECHN as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check INNOVATIVE STERILIZATION TECHN yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Innovative Sterilization Technologies LLC
Location
Dayton, OH (US)
Source
device