BREAS MEDICAL, INC. FDA Inspection Record
BREAS MEDICAL, INC. is an FDA-registered facility (FEI 3010817335) located in N Billerica, MA, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 18, 2022: No Action Indicated (NAI)
- Inspections on record
- 4 (3 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Jul 25, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 18, 2022 | No Action Indicated (NAI) | Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) |
| Feb 6, 2020 | Voluntary Action Indicated (VAI) | DHR - not or inadequately maintained (21 CFR 820.184) Lack of or inadequate complaint procedures (21 CFR 820.198(a)) |
| Nov 25, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Oct 2, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 25, 2024 | Class I | Z-2690-2024 | Vivo 45 LS, pressure and volume ventilator capable of delivering continuous or… There is a potential for short term (<7 days) elevated levels of formaldehyde emitted from the device into the patient breathing airpath in… |
| Oct 13, 2023 | Class II | Z-0405-2024 | with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653… Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing… |
| Jun 2, 2022 | Class II | Z-1382-2022 | Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support… A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor… |
| Mar 21, 2022 | Class II | Z-0943-2022 | Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models
Article… Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley… |
| Jan 28, 2020 | Class II | Z-1352-2020 | Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory… The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover. |
FDA record score
19 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-2690-2024, this FEI as the recalling firm | Jul 25, 2024 | 15 x 0.6 | 9 |
| Class II recall Z-0405-2024, this FEI as the recalling firm | Oct 13, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1382-2022, this FEI as the recalling firm | Jun 2, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0943-2022, this FEI as the recalling firm | Mar 21, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Feb 6, 2020 | 6 x 0.25 | 1.5 |
| Class II recall Z-1352-2020, this FEI as the recalling firm | Jan 28, 2020 | 4 x 0.25 | 1 |
| Inspection classified VAI | Nov 25, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 2, 2015 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing BREAS MEDICAL, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check BREAS MEDICAL, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Human Design Medical, LLC
- Location
- N Billerica, MA (US)
- Source
- device