Microgenics Corporation FDA Inspection Record
Microgenics Corporation is an FDA-registered facility (FEI 3010939897) located in Fremont, CA, US. RegDataLab tracks 9 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Dec 15, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 9 (3 VAI, 6 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 6, latest Aug 17, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Dec 15, 2016 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) |
| Jul 16, 2015 | No Action Indicated (NAI) | No citations listed |
| Oct 20, 2014 | No Action Indicated (NAI) | Design history file (21 CFR 820.30(j)) |
| Oct 25, 2013 | No Action Indicated (NAI) | No citations listed |
| Oct 25, 2013 | No Action Indicated (NAI) | No citations listed |
| Jun 14, 2012 | Voluntary Action Indicated (VAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Investigation of device failures (21 CFR 820.198(c)) |
| Jun 14, 2012 | Voluntary Action Indicated (VAI) | No citations listed |
| Jun 2, 2010 | No Action Indicated (NAI) | No citations listed |
| Jan 28, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 17, 2026 | Class II | Z-0159-2027 | MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF… Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low… |
| Oct 1, 2025 | Class II | Z-0415-2026 | Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control… The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I… |
| Jan 20, 2022 | Class II | Z-1284-2022 | QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET
QMS TACROLIMUS CAL A… Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient… |
| Aug 15, 2016 | Class II | Z-0008-2017 | Thermo Scientific QMS Everolimus Assay;
Model number:
0380000 (US… A reagent stability issue (still under investigation) causes less separation in the calibration curve, which affects results across the… |
| Jul 25, 2016 | Class II | Z-2884-2016 | DRI Salicylate Serum Tox Assay, Catalog number 0977.
Toxicology:
The DRI Salicylate… Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL. |
| Oct 20, 2014 | Class II | Z-0501-2015 | MAS CardioImmune XL Cardiac Marker Control;
Model: CAI-XL1;
intended for use in the… MAS CardioImmune XL, level 1 (lot CXL16011), is showing vial-to-vial variation for the analytes Creatine Kinase-MB (CK-MB) and B-Type… |
FDA record score
12 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0159-2027, this FEI as the recalling firm | Aug 17, 2026 | 4 x 1 | 4 |
| Class II recall Z-0415-2026, this FEI as the recalling firm | Oct 1, 2025 | 4 x 1 | 4 |
| Class II recall Z-1284-2022, this FEI as the recalling firm | Jan 20, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Dec 15, 2016 | 6 x 0.25 | 1.5 |
| Class II recall Z-0008-2017, this FEI as the recalling firm | Aug 15, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2884-2016, this FEI as the recalling firm | Jul 25, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0501-2015, this FEI as the recalling firm | Oct 20, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 14, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 14, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Microgenics Corporation as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Microgenics Corporation yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Microgenics Corporation
- Location
- Fremont, CA (US)
- Source