Endologix LLC FDA Inspection Record
Endologix LLC is an FDA-registered facility (FEI 3011063223) located in Irvine, CA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Aug 3, 2021: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (1 OAI, 2 VAI, 2 NAI)
- Last OAI inspection
- Feb 20, 2020
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 8, latest Jul 31, 2018
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 3, 2021 | Voluntary Action Indicated (VAI) | Design input - documentation (21 CFR 820.30(c)) Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) and 4 more |
| Feb 20, 2020 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Nonconforming product control (21 CFR 820.90(a)) and 2 more |
| May 20, 2016 | No Action Indicated (NAI) | No citations listed |
| Jan 22, 2015 | No Action Indicated (NAI) | No citations listed |
| Sep 25, 2014 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 31, 2018 | Class I | Z-0006-2019 | ENDOLOGIX AFX Endovascular AAA System, Stent Graft System, Rx Only, Sterile EO, REF… and 3 more products Endologix is notifying physicians of patient tailored surveillance recommendations as well as general warnings/precautions for interventions… |
| Jan 4, 2017 | Class II | Z-1048-2017 | AFX Endovascular AAA System
Item Number Item Description
F0070305… and 1 more product Endologix updated information on the rates of Type III endoleaks and suggestions for patient surveillance and treatment. |
| Dec 27, 2016 | Class II | Z-1035-2017 | AFX Endovascular AAA System, Endoleak Type IIIA Endologix initiated a correction by updating reports for the cumulative clinical data and information on the Powerlink System and the AFX… |
| Dec 27, 2016 | Class II | Z-1037-2017 | AFX Endovascular AAA System, Endoleak Type IIIA
Common Name: AFX Bifurcated and… Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System). |
| Dec 27, 2016 | Class II | Z-1038-2017 | AFX Endovascular AAA System, Endoleak Type IIIB;
The AFX Endovascular AAA System is… Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System). |
| Dec 27, 2016 | Class II | Z-1039-2017 | AFX Endovascular AAA System, Endoleak Type: IIIB
Common Name: AFX Bifurcated and… Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System). |
| Dec 27, 2016 | Class II | Z-1047-2017 | AFX Endovascular AAA System, Endoleak Type IIIA
Product Usage:
The AFX Endovascular… Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System). |
| Dec 27, 2016 | Class II | Z-1054-2017 | AFX Endovascular AAA System, Endoleak Type IIIB
Common Name: AFX Bifurcated and… Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System). |
FDA record score
20 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
11 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified OAI | Feb 20, 2020 | 30 x 0.25 | 7.5 |
| Class I recall Z-0006-2019, this FEI as the recalling firm | Jul 31, 2018 | 15 x 0.25 | 3.8 |
| Inspection classified VAI | Aug 3, 2021 | 6 x 0.25 | 1.5 |
| Class II recall Z-1048-2017, this FEI as the recalling firm | Jan 4, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1035-2017, this FEI as the recalling firm | Dec 27, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-1037-2017, this FEI as the recalling firm | Dec 27, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-1038-2017, this FEI as the recalling firm | Dec 27, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-1039-2017, this FEI as the recalling firm | Dec 27, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-1047-2017, this FEI as the recalling firm | Dec 27, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-1054-2017, this FEI as the recalling firm | Dec 27, 2016 | 4 x 0.25 | 1 |
| Inspection classified VAI | Sep 25, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Endologix LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Endologix LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- Endologix LLC
- Location
- Irvine, CA (US)
- Source
- device