NEUROMETRIX, INC. FDA Inspection Record
NEUROMETRIX, INC. is an FDA-registered facility (FEI 3011412151) located in Rockaway, NJ, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 27, 2026: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (5 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 27, 2026 | Voluntary Action Indicated (VAI) | Records of risk management activities in product realization not maintained (ISO 13485:2016 Clause 7.1) Design and development verification not performed (ISO 13485:2016 Clause 7.3.6) Lack of documented justification for not investigating complaint(s) (ISO 13485:2016 Clause 8.2.2) Lack of Written MDR Procedures (21 CFR 803.17) |
| Nov 15, 2017 | No Action Indicated (NAI) | No citations listed |
| May 7, 2015 | Voluntary Action Indicated (VAI) | ID label, labeling (21 CFR 820.184(e)) |
| May 7, 2015 | Voluntary Action Indicated (VAI) | Incorrect translation to production specifications (21 CFR 820.30(h)) Training records (21 CFR 820.25(b)) |
| Mar 20, 2012 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) DHR content (21 CFR 820.184) |
| Apr 23, 2009 | Voluntary Action Indicated (VAI) | Maintained (21 CFR 820.198(e)) Lack of Written MDR Procedures (21 CFR 803.17) |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | May 27, 2026 | 6 x 1 | 6 |
| Inspection classified VAI | May 7, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 7, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 20, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 23, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |