ATRICURE, INC. FDA Inspection Record

FEI 3011706110 · Inspections, Form 483s, warning letters and import refusals

ATRICURE, INC. is an FDA-registered facility (FEI 3011706110) located in Mason, OH, US. RegDataLab tracks 10 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Dec 13, 2018: No Action Indicated (NAI)
Inspections on record
10 (5 VAI, 5 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
6, latest Nov 28, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 13, 2018No Action Indicated (NAI)No citations listed
Mar 9, 2018Voluntary Action Indicated (VAI)No citations listed
Mar 9, 2018Voluntary Action Indicated (VAI)Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b))
Documentation (21 CFR 820.80(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
Oct 13, 2016No Action Indicated (NAI)No citations listed
Aug 29, 2016Voluntary Action Indicated (VAI)Design validation - Risk analysis (21 CFR 820.30(g))
Lack of Written MDR Procedures (21 CFR 803.17)
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Feb 26, 2013No Action Indicated (NAI)No citations listed
Jan 18, 2013Voluntary Action Indicated (VAI)Report of risk to health (21 CFR 806.10(a)(1))
Apr 1, 2011No Action Indicated (NAI)No citations listed
Feb 22, 2011Voluntary Action Indicated (VAI)No citations listed
Jun 21, 2010No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Nov 28, 2023Class IIZ-0927-2024Cryo Module Accessories Domestic CMA-OUS, Product Catalog Number A001053 and 1 more product
AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These…
May 29, 2019Class IIZ-1887-2019EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The…
The sterile package seal for the Epi-Sense Guided Coagulation System may be compromised, and has a worst-case reasonable harm of infection…
Jan 22, 2019Class IIZ-0829-2019COBRA FUSION 50, Ablation System, Catalog # (REF) 700-002 and 3 more products
There is a potential for patients undergoing ablation to experience an elevated risk for TE; during stand-alone off-pump procedures where…
Nov 30, 2016Class IIZ-1148-2017AtriClip" LAA Exclusion System with preloaded Gillinov-Cosgrove" Clip - A single PRO2…
A complaint of the PRO2 jaw breaking prior to surgery being performed.
Sep 22, 2016Class IIZ-0377-2017AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip Product is used for…
There is a potential for a component failure of the device which could necessitate medical intervention.
Sep 22, 2016Class IIZ-0653-2017COBRA Fusion 50 Ablation System, COBRA Fusion 150 Ablation System, and COBRA Fusion…
Due to product design and process control, part of the device can separate and remain behind when the device is removed that can potentially…

FDA record score

8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

11 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0927-2024, this FEI as the recalling firmNov 28, 20234 x 0.62.4
Inspection classified VAIMar 9, 20186 x 0.251.5
Inspection classified VAIMar 9, 20186 x 0.251.5
Class II recall Z-1887-2019, this FEI as the recalling firmMay 29, 20194 x 0.251
Class II recall Z-0829-2019, this FEI as the recalling firmJan 22, 20194 x 0.251
Class II recall Z-1148-2017, this FEI as the recalling firmNov 30, 20164 x 0.251
Class II recall Z-0377-2017, this FEI as the recalling firmSep 22, 201610 years or older, so it adds no points0
Class II recall Z-0653-2017, this FEI as the recalling firmSep 22, 201610 years or older, so it adds no points0
Inspection classified VAIAug 29, 201610 years or older, so it adds no points0
Inspection classified VAIJan 18, 201310 years or older, so it adds no points0
Inspection classified VAIFeb 22, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing ATRICURE, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check ATRICURE, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
AtriCure, Inc.
Location
Mason, OH (US)
Source
device