ConvaTec, Inc. FDA Inspection Record

FEI 3011987967 · Inspections, Form 483s, warning letters and import refusals

ConvaTec, Inc. is an FDA-registered facility (FEI 3011987967) located in Greensboro, NC, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Sep 8, 2015: No Action Indicated (NAI)
Inspections on record
6 (2 OAI, 1 VAI, 3 NAI)
Last OAI inspection
Apr 16, 2014
Warning letters
1 listed by FDA for this FEI, latest May 24, 2013
Recalls as the recalling firm
20, latest Jul 6, 2026
Import refusals
10 or more

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 8, 2015No Action Indicated (NAI)No citations listed
Apr 16, 2014Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Design validation - user needs and intended uses (21 CFR 820.30(g))
and 5 more
Sep 11, 2013No Action Indicated (NAI)No citations listed
Feb 5, 2013Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
Lack of Written MDR Procedures (21 CFR 803.17)
and 2 more
Mar 23, 2011No Action Indicated (NAI)No citations listed
Jul 1, 2010Voluntary Action Indicated (VAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Records of MDR Investigation (21 CFR 820.198(d))
CAPA verification/validation of corrective/preventive action (21 CFR 820.100(a)(4))
Report of Death or Serious Injury (21 CFR 803.50(a)(1))

Recalls

Showing the 10 most recent of 20 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 6, 2026Class IIZ-2986-2026Esteem Body Soft Convex Urostomy Pouch Model/Catalog Numbers: 423703, 423704, 423705…
Urostomy Pouches may leak from cap during use.
Jun 8, 2026Class IIZ-2677-2026Aquacel Ag Surgical SP REF: 413556 9cm x 15cm/3.5 in. x 6 in., Sterile R
Surgical dressing manufactured out of specification (greater thickness).
Feb 16, 2026Class IIZ-1582-2026Convatec, EsteemBody Drainable Pouch REF:423643, 10-45mm and 2 more products
Drainable large pouch may leak due to manufacturing issue.
Aug 29, 2025Class IIZ-0160-2026DuoDERM" Extra Thin dressings are highly flexible, control gel formula dressings designed…
Wound dressing may have foreign matter on the product.
Nov 29, 2024Class IIZ-0903-2025Esteem Synergy Stomahesive Skin Barrier, part of Esteem synergy, which is a two-piece…
Ostomy skin barrier package contains a pre-cut wafer of 28mm instead of 16mm and/or 35mm pre-cut wafers, as stated package labeling. If…
Feb 27, 2024Class IIZ-1495-2024ConvaTec CarboFlex, 15cm x 20cm, 6 in. x 8 in., Order No. 403204, Odor Control Dressing…
The product does not meet sterility assurance level.
Feb 19, 2024Class IIZ-1350-2024Convatec EsteemBody Soft Convex, Drainable Pouch, Product codes (REF): a) 423643, b)…
Convatec Inc is conducting a Voluntary Medical Device Recall (removal) for specific lots of the EsteemBody Soft Convex, Drainable Pouch due…
Jan 17, 2024Class IIZ-1276-2024ConvaTec Natura Stomahesive Post-Operative / Surgical Ostomy System, REF 416938, 1 3/4 in…
The kit contains a flange that has a 45mm coupling ring, however, the pouch within the kit is for a 57mm coupling ring.
Nov 15, 2022Class IIZ-0723-2023ConvaTec AQUACEL Foam Ag Adhesive, wound dressing advesive
ConvaTec are voluntarily recalling a batch of AQUACEL¿ Foam Ag Adhesive 10cm x 10cm (x 10 pack) due to the identification of visual…
Aug 16, 2022Class IIZ-0127-2023ConvaTec Esteem+ Durahesive Plus, Drainable Pouch, 3/4-2 1/2 in., 19-64mm, REF 416976. For… and 3 more products
impaired functionality of the release liner

FDA record score

42 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

24 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2986-2026, this FEI as the recalling firmJul 6, 20264 x 14
Class II recall Z-2677-2026, this FEI as the recalling firmJun 8, 20264 x 14
Class II recall Z-1582-2026, this FEI as the recalling firmFeb 16, 20264 x 14
Class II recall Z-0160-2026, this FEI as the recalling firmAug 29, 20254 x 14
Class II recall Z-0903-2025, this FEI as the recalling firmNov 29, 20244 x 14
Class II recall Z-1495-2024, this FEI as the recalling firmFeb 27, 20244 x 0.62.4
Class II recall Z-1350-2024, this FEI as the recalling firmFeb 19, 20244 x 0.62.4
Class II recall Z-1276-2024, this FEI as the recalling firmJan 17, 20244 x 0.62.4
Class II recall Z-0723-2023, this FEI as the recalling firmNov 15, 20224 x 0.62.4
Class II recall Z-0127-2023, this FEI as the recalling firmAug 16, 20224 x 0.62.4
Class II recall Z-1134-2022, this FEI as the recalling firmApr 22, 20224 x 0.62.4
Class II recall Z-2614-2021, this FEI as the recalling firmAug 20, 20214 x 0.251
Class II recall Z-2519-2021, this FEI as the recalling firmAug 9, 20214 x 0.251
Class II recall Z-0233-2020, this FEI as the recalling firmSep 6, 20194 x 0.251
Class II recall Z-1035-2019, this FEI as the recalling firmJan 9, 20194 x 0.251
Class II recall Z-2744-2018, this FEI as the recalling firmJul 3, 20184 x 0.251
Class II recall Z-1890-2018, this FEI as the recalling firmApr 25, 20184 x 0.251
Class II recall Z-0525-2018, this FEI as the recalling firmOct 26, 20174 x 0.251
Class II recall Z-2228-2017, this FEI as the recalling firmMay 10, 20174 x 0.251
Class II recall Z-1044-2016, this FEI as the recalling firmNov 11, 201510 years or older, so it adds no points0
Inspection classified OAIApr 16, 201410 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsMay 24, 201310 years or older, so it adds no points0
Inspection classified OAIFeb 5, 201310 years or older, so it adds no points0
Inspection classified VAIJul 1, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing ConvaTec, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check ConvaTec, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
ConvaTec, Inc
Location
Greensboro, NC (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.