ConvaTec, Inc. FDA Inspection Record
ConvaTec, Inc. is an FDA-registered facility (FEI 3011987967) located in Greensboro, NC, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Sep 8, 2015: No Action Indicated (NAI)
- Inspections on record
- 6 (2 OAI, 1 VAI, 3 NAI)
- Last OAI inspection
- Apr 16, 2014
- Warning letters
- 1 listed by FDA for this FEI, latest May 24, 2013
- Recalls as the recalling firm
- 20, latest Jul 6, 2026
- Import refusals
- 10 or more
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 8, 2015 | No Action Indicated (NAI) | No citations listed |
| Apr 16, 2014 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Design validation - user needs and intended uses (21 CFR 820.30(g)) and 5 more |
| Sep 11, 2013 | No Action Indicated (NAI) | No citations listed |
| Feb 5, 2013 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of Written MDR Procedures (21 CFR 803.17) and 2 more |
| Mar 23, 2011 | No Action Indicated (NAI) | No citations listed |
| Jul 1, 2010 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Records of MDR Investigation (21 CFR 820.198(d)) CAPA verification/validation of corrective/preventive action (21 CFR 820.100(a)(4)) Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 6, 2026 | Class II | Z-2986-2026 | Esteem Body Soft Convex Urostomy Pouch
Model/Catalog Numbers: 423703, 423704, 423705… Urostomy Pouches may leak from cap during use. |
| Jun 8, 2026 | Class II | Z-2677-2026 | Aquacel Ag Surgical SP
REF: 413556
9cm x 15cm/3.5 in. x 6 in., Sterile R Surgical dressing manufactured out of specification (greater thickness). |
| Feb 16, 2026 | Class II | Z-1582-2026 | Convatec, EsteemBody Drainable Pouch
REF:423643,
10-45mm and 2 more products Drainable large pouch may leak due to manufacturing issue. |
| Aug 29, 2025 | Class II | Z-0160-2026 | DuoDERM" Extra Thin dressings are highly flexible, control gel formula dressings designed… Wound dressing may have foreign matter on the product. |
| Nov 29, 2024 | Class II | Z-0903-2025 | Esteem Synergy Stomahesive Skin Barrier, part of Esteem synergy, which is a two-piece… Ostomy skin barrier package contains a pre-cut wafer of 28mm instead of 16mm and/or 35mm pre-cut wafers, as stated package labeling. If… |
| Feb 27, 2024 | Class II | Z-1495-2024 | ConvaTec CarboFlex, 15cm x 20cm, 6 in. x 8 in., Order No. 403204, Odor Control Dressing… The product does not meet sterility assurance level. |
| Feb 19, 2024 | Class II | Z-1350-2024 | Convatec EsteemBody Soft Convex, Drainable Pouch, Product codes (REF): a) 423643, b)… Convatec Inc is conducting a Voluntary Medical Device Recall (removal) for specific lots of the EsteemBody Soft Convex, Drainable Pouch due… |
| Jan 17, 2024 | Class II | Z-1276-2024 | ConvaTec Natura Stomahesive Post-Operative / Surgical Ostomy System, REF 416938, 1 3/4 in… The kit contains a flange that has a 45mm coupling ring, however, the pouch within the kit is for a 57mm coupling ring. |
| Nov 15, 2022 | Class II | Z-0723-2023 | ConvaTec AQUACEL Foam Ag Adhesive, wound dressing advesive ConvaTec are voluntarily recalling a batch of AQUACEL¿ Foam Ag Adhesive 10cm x 10cm (x 10 pack) due to the identification of visual… |
| Aug 16, 2022 | Class II | Z-0127-2023 | ConvaTec Esteem+ Durahesive Plus, Drainable Pouch, 3/4-2 1/2 in., 19-64mm, REF 416976. For… and 3 more products impaired functionality of the release liner |
FDA record score
42 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
24 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2986-2026, this FEI as the recalling firm | Jul 6, 2026 | 4 x 1 | 4 |
| Class II recall Z-2677-2026, this FEI as the recalling firm | Jun 8, 2026 | 4 x 1 | 4 |
| Class II recall Z-1582-2026, this FEI as the recalling firm | Feb 16, 2026 | 4 x 1 | 4 |
| Class II recall Z-0160-2026, this FEI as the recalling firm | Aug 29, 2025 | 4 x 1 | 4 |
| Class II recall Z-0903-2025, this FEI as the recalling firm | Nov 29, 2024 | 4 x 1 | 4 |
| Class II recall Z-1495-2024, this FEI as the recalling firm | Feb 27, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1350-2024, this FEI as the recalling firm | Feb 19, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1276-2024, this FEI as the recalling firm | Jan 17, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0723-2023, this FEI as the recalling firm | Nov 15, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0127-2023, this FEI as the recalling firm | Aug 16, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1134-2022, this FEI as the recalling firm | Apr 22, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-2614-2021, this FEI as the recalling firm | Aug 20, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2519-2021, this FEI as the recalling firm | Aug 9, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0233-2020, this FEI as the recalling firm | Sep 6, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1035-2019, this FEI as the recalling firm | Jan 9, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2744-2018, this FEI as the recalling firm | Jul 3, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1890-2018, this FEI as the recalling firm | Apr 25, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0525-2018, this FEI as the recalling firm | Oct 26, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2228-2017, this FEI as the recalling firm | May 10, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1044-2016, this FEI as the recalling firm | Nov 11, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 16, 2014 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 24, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Feb 5, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 1, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing ConvaTec, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ConvaTec, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- ConvaTec, Inc
- Location
- Greensboro, NC (US)
- Source