TURBETT SURGICAL, Inc FDA Inspection Record
TURBETT SURGICAL, Inc is an FDA-registered facility (FEI 3012421755) located in Macedon, NY, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Nov 6, 2025: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (4 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Dec 5, 2022
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 6, 2025 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Statistical techniques - Lack of or inadequate procedures (21 CFR 820.250(a)) Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2)) Documentation (21 CFR 820.100(b)) |
| Jun 26, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) and 5 more |
| Jun 28, 2022 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 4 more |
| Mar 6, 2019 | Voluntary Action Indicated (VAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Design output - Lack of or inadequate procedures (21 CFR 820.30(d)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Dec 5, 2022 | Class II | Z-1037-2023 | Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are… and 1 more product Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major… |
| Apr 14, 2022 | Class II | Z-1501-2022 | Turbett Surgical Container, TS1500 Devices were subject to unapproved rework processes. |
| Apr 1, 2022 | Class II | Z-1111-2022 | Turbett Surgical Container, TS1500 Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after prolonged use. |
FDA record score
20 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Nov 6, 2025 | 6 x 1 | 6 |
| Inspection classified VAI | Jun 26, 2023 | 6 x 0.6 | 3.6 |
| Inspection classified VAI | Jun 28, 2022 | 6 x 0.6 | 3.6 |
| Class II recall Z-1037-2023, this FEI as the recalling firm | Dec 5, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1501-2022, this FEI as the recalling firm | Apr 14, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1111-2022, this FEI as the recalling firm | Apr 1, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Mar 6, 2019 | Only the 3 most recent VAI inspections add points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 27, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing TURBETT SURGICAL, Inc as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for TURBETT SURGICAL, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about TURBETT SURGICAL, Inc → Check this supplier
Facility details
- Company
- Turbett Surgical, Inc.
- Location
- Macedon, NY (US)
- Source
- device