TURBETT SURGICAL, Inc FDA Inspection Record

FEI 3012421755 · Inspections, Form 483s, warning letters and import refusals

TURBETT SURGICAL, Inc is an FDA-registered facility (FEI 3012421755) located in Macedon, NY, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Nov 6, 2025: Voluntary Action Indicated (VAI)
Inspections on record
4 (4 VAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
3, latest Dec 5, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Nov 6, 2025Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Statistical techniques - Lack of or inadequate procedures (21 CFR 820.250(a))
Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2))
Documentation (21 CFR 820.100(b))
Jun 26, 2023Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h))
and 5 more
Jun 28, 2022Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 4 more
Mar 6, 2019Voluntary Action Indicated (VAI)Design input - Lack of or inadequate procedures (21 CFR 820.30(c))
Design output - Lack of or inadequate procedures (21 CFR 820.30(d))
Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h))
Lack of or inadequate procedures (21 CFR 820.100(a))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 5, 2022Class IIZ-1037-2023Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are… and 1 more product
Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major…
Apr 14, 2022Class IIZ-1501-2022Turbett Surgical Container, TS1500
Devices were subject to unapproved rework processes.
Apr 1, 2022Class IIZ-1111-2022Turbett Surgical Container, TS1500
Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after prolonged use.

FDA record score

20 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 27, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAINov 6, 20256 x 16
Inspection classified VAIJun 26, 20236 x 0.63.6
Inspection classified VAIJun 28, 20226 x 0.63.6
Class II recall Z-1037-2023, this FEI as the recalling firmDec 5, 20224 x 0.62.4
Class II recall Z-1501-2022, this FEI as the recalling firmApr 14, 20224 x 0.62.4
Class II recall Z-1111-2022, this FEI as the recalling firmApr 1, 20224 x 0.62.4
Inspection classified VAIMar 6, 2019Only the 3 most recent VAI inspections add points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing TURBETT SURGICAL, Inc as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for TURBETT SURGICAL, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about TURBETT SURGICAL, Inc → Check this supplier

Facility details

Company
Turbett Surgical, Inc.
Location
Macedon, NY (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.