Getinge USA Sales Inc FDA Inspection Record

FEI 3013876692 · Inspections, Form 483s, warning letters and import refusals

Getinge USA Sales Inc (FEI 3013876692) located in Wayne, New Jersey, United States has FDA import refusal records.

What FDA found at this site

Inspections on record
None in FDA's published inspection data
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
28, latest Jan 15, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Recalls

Showing the 10 most recent of 28 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jan 15, 2025Class IZ-1074-2025Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i…
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers…
Oct 3, 2024Class IIZ-0456-2025MEERA CL mobile operating table - Designed for the placement and positioning of the… and 7 more products
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur…
May 29, 2024Class IIZ-2271-2024ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322B4 and 2 more products
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile…
May 1, 2024Class IZ-1877-2024Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of…
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers…
Dec 11, 2023Class IIZ-0987-2024Flow-i C20 Anesthesia System, model 6677200 and 4 more products
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote…
Nov 8, 2023Class IIZ-0757-2024Maquet PowerLEDII OR Light System, Model Names and Catalog Numbers PWDII77SF… and 14 more products
A potential for a light system to fall in the operating room.
Sep 6, 2023Class IIZ-0036-2024Meera Mobile Operating Table-to support and position the patient immediately before…
Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit…
Dec 22, 2022Class IZ-0959-2023Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no… and 2 more products
Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a…
Nov 7, 2022Class IIZ-0422-2023Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code…
The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the…
Oct 14, 2022Class IIZ-0239-2023Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first…

FDA record score

Not scored. FDA's published inspection data has no inspection for this site, and the score is only counted for inspected sites. This is not a score of 0. The records FDA does list for this FEI are below.

Counted on Nov 29, 2026, method version 1. How this is counted.

26 records FDA lists for this FEI
RecordDatePoints x age weightAdds
Class I recall Z-1074-2025, this FEI as the recalling firmJan 15, 2025The site is not scored, so it adds no points0
Class I recall Z-1877-2024, this FEI as the recalling firmMay 1, 2024The site is not scored, so it adds no points0
Class I recall Z-0959-2023, this FEI as the recalling firmDec 22, 2022The site is not scored, so it adds no points0
Class I recall Z-1560-2022, this FEI as the recalling firmJul 20, 2022The site is not scored, so it adds no points0
Class I recall Z-1278-2022, this FEI as the recalling firmMay 2, 2022The site is not scored, so it adds no points0
Class I recall Z-0468-2022, this FEI as the recalling firmDec 9, 2021The site is not scored, so it adds no points0
Class II recall Z-0456-2025, this FEI as the recalling firmOct 3, 2024The site is not scored, so it adds no points0
Class II recall Z-2271-2024, this FEI as the recalling firmMay 29, 2024The site is not scored, so it adds no points0
Class II recall Z-0987-2024, this FEI as the recalling firmDec 11, 2023The site is not scored, so it adds no points0
Class II recall Z-0757-2024, this FEI as the recalling firmNov 8, 2023The site is not scored, so it adds no points0
Class II recall Z-0036-2024, this FEI as the recalling firmSep 6, 2023The site is not scored, so it adds no points0
Class II recall Z-0422-2023, this FEI as the recalling firmNov 7, 2022The site is not scored, so it adds no points0
Class II recall Z-0239-2023, this FEI as the recalling firmOct 14, 2022The site is not scored, so it adds no points0
Class II recall Z-1568-2022, this FEI as the recalling firmJul 21, 2022The site is not scored, so it adds no points0
Class II recall Z-1308-2022, this FEI as the recalling firmMay 6, 2022The site is not scored, so it adds no points0
Class II recall Z-1248-2022, this FEI as the recalling firmApr 13, 2022The site is not scored, so it adds no points0
Class II recall Z-1078-2022, this FEI as the recalling firmApr 1, 2022The site is not scored, so it adds no points0
Class II recall Z-1181-2021, this FEI as the recalling firmJan 29, 2021The site is not scored, so it adds no points0
Class II recall Z-0677-2020, this FEI as the recalling firmNov 13, 2019The site is not scored, so it adds no points0
Class II recall Z-2559-2019, this FEI as the recalling firmAug 22, 2019The site is not scored, so it adds no points0
Class II recall Z-2433-2019, this FEI as the recalling firmJul 22, 2019The site is not scored, so it adds no points0
Class II recall Z-2077-2019, this FEI as the recalling firmJun 5, 2019The site is not scored, so it adds no points0
Class II recall Z-0987-2019, this FEI as the recalling firmJan 8, 2019The site is not scored, so it adds no points0
Class II recall Z-1349-2018, this FEI as the recalling firmFeb 7, 2018The site is not scored, so it adds no points0
Class II recall Z-1172-2018, this FEI as the recalling firmJan 3, 2018The site is not scored, so it adds no points0
Class II recall Z-1927-2018, this FEI as the recalling firmNov 16, 2017The site is not scored, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 29, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 29, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 29, 2026Newest record Nov 19, 2026Current
Warning letter textNov 29, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 28, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 29, 2026Newest record Nov 19, 2026Current

Reviewing Getinge USA Sales Inc as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Getinge USA Sales Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Getinge USA Sales Inc → Check this supplier

Facility details

FDA Legal Name
Getinge Usa Sales Inc
Location
Wayne, New Jersey (United States)

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.