ISOPURE, CORP. — FDA Facility Profile
ISOPURE, CORP. is an FDA-registered facility (FEI 3013942511) located in Louisville, KY, US. RegDataLab tracks 5 FDA inspections at this facility.
Reviewing ISOPURE, CORP. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ISOPURE, CORP. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
What FDA found at this site
- Latest inspection
- Aug 15, 2024: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (4 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Jul 26, 2022
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 15, 2024 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) |
| Oct 21, 2022 | No Action Indicated (NAI) | No citations listed |
| Sep 23, 2019 | Voluntary Action Indicated (VAI) | Records of complaint investigation (21 CFR 820.198(e)) Time to report - 10 days (21 CFR 806.10(b)) |
| Sep 28, 2017 | Voluntary Action Indicated (VAI) | Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Aug 25, 2011 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) Report of risk to health (21 CFR 806.10(a)(1)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 26, 2022 | Not yet classified | Z-1739-2022 | Acidified Concentrate Distribution and Storage-The Isopure ACDS (Acidified Concentrate… Acidified Concentrate Distribution System potential electrical safety hazard may result in electric shock, arching |
| Apr 28, 2022 | Not yet classified | Z-1360-2022 | Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) Isopure received a report of a hazardous incident involving a device which had experienced an electrical short that resulted in a thermal… |
| Apr 19, 2019 | Not yet classified | Z-0821-2020 | Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson… and 2 more products There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This… |
Facility details
- Company
- Isopure Corp.
- Location
- Louisville, KY (US)
- Source