XCISION MEDICAL SYSTEMS, LLC — FDA Facility Profile

FEI 3014273653

XCISION MEDICAL SYSTEMS, LLC is an FDA-registered facility (FEI 3014273653) located in Columbia, MD, US. RegDataLab tracks 1 FDA inspection at this facility.

Reviewing XCISION MEDICAL SYSTEMS, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check XCISION MEDICAL SYSTEMS, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

What FDA found at this site

Latest inspection
Jan 9, 2020: No Action Indicated (NAI)
Inspections on record
1 (1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
2, latest Nov 30, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jan 9, 2020No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Nov 30, 2023Not yet classifiedZ-0750-2024GammaPod - Treatment Planning System-a component of the GammaPod Irradiation System…
Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be…
Jul 3, 2019Not yet classifiedZ-2081-2019GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product…
One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

Facility details

Company
Xcision Medical Systems, LLC
Location
Columbia, MD (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.