Cardinal Health FDA Inspection Record
Cardinal Health (FEI 3014652171) located in Mansfield, Massachusetts, United States has FDA import refusal records.
What FDA found at this site
- Inspections on record
- None in FDA's published inspection data
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 11, latest Aug 4, 2021
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 4, 2021 | Class I | Z-2284-2021 | Monoject 0.9% Sodium Chloride Flush Syringe, 10 mL Fill, STERILE, Product Code… and 2 more products Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe. |
| Jun 16, 2021 | Class I | Z-2244-2021 | Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number… The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a… |
| Jun 9, 2021 | Class II | Z-2094-2021 | Infant Heel Warmer with Attachment Tab, Nonsterile, Active Product
SKU MH00002T
The… and 1 more product A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during… |
| Feb 9, 2021 | Class II | Z-1250-2021 | Magellan Hypodermic Safety Needle, 25 G x 5/8" (0.508 mm x 1.6 cm). A single lumen needle… and 2 more products Safety shield not locking into the cannula. If the safety shield activation fails to lock a contaminated needle, a needlestick may occur and… |
| Sep 25, 2020 | Class II | Z-0424-2021 | Covidien MONOJECT Standard Hypodermic Needle, Polypropylene Hub 22 G x 3/4". Intended use… Potential for the needle to detach from the cartridge after the bottom of the cartridge is twisted and removed. If the needle is not… |
| Sep 22, 2020 | Class II | Z-0452-2021 | Kangaroo Connect 500ml Bag Set, Non-Sterile, Product Code 77500FD - Product Usage… and 1 more product The feeding spike sets may leak at the interface of the tube and spike connector. |
| Jan 15, 2020 | Class II | Z-1453-2020 | Monoject 3mL Syringe with Hypodermic Safety Needle, 22G x 1-1/2 . Item Code 11832215… and 3 more products Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo… |
| Oct 28, 2019 | Class II | Z-0990-2020 | Genius 2 Tympanic Thermometer (discontinued)
Item Code: 303000 and 4 more products The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always… |
| Apr 30, 2019 | Class II | Z-2036-2019 | Monoject Standard Hypodermic Needle, 14 G x 1-1/2"
Item code: 8881200011 and 16 more products Manufacturing defect was found for the cartridge component, which could compromise the sterility barrier of the product |
| Mar 22, 2019 | Class II | Z-1283-2019 | ChemoPlus Chemo Spill Kit
Product Code:CT4004
Product Usage:
ChemoPlus chemotherapy… and 3 more products Kit label states not made with natural rubber latex,however safety goggles within the kits contain natural rubber latex in the goggle strap |
FDA record score
Not scored. FDA's published inspection data has no inspection for this site, and the score is only counted for inspected sites. This is not a score of 0. The records FDA does list for this FEI are below.
11 records FDA lists for this FEI
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-2284-2021, this FEI as the recalling firm | Aug 4, 2021 | The site is not scored, so it adds no points | 0 |
| Class I recall Z-2244-2021, this FEI as the recalling firm | Jun 16, 2021 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-2094-2021, this FEI as the recalling firm | Jun 9, 2021 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-1250-2021, this FEI as the recalling firm | Feb 9, 2021 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-0424-2021, this FEI as the recalling firm | Sep 25, 2020 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-0452-2021, this FEI as the recalling firm | Sep 22, 2020 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-1453-2020, this FEI as the recalling firm | Jan 15, 2020 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-0990-2020, this FEI as the recalling firm | Oct 28, 2019 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-2036-2019, this FEI as the recalling firm | Apr 30, 2019 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-1283-2019, this FEI as the recalling firm | Mar 22, 2019 | The site is not scored, so it adds no points | 0 |
| Class II recall Z-0957-2019, this FEI as the recalling firm | Dec 27, 2018 | The site is not scored, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 29, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 29, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 29, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
Reviewing Cardinal Health as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Cardinal Health: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Location
- Mansfield, Massachusetts (United States)