Murata Vios, Inc. FDA Inspection Record

FEI 3015328411 · Inspections, Form 483s, warning letters and import refusals

Murata Vios, Inc. is an FDA-registered facility (FEI 3015328411) located in Woodbury, MN, US. RegDataLab tracks 2 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Dec 11, 2024: Voluntary Action Indicated (VAI)
Inspections on record
2 (1 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
4, latest Aug 9, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Dec 11, 2024Voluntary Action Indicated (VAI)Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Lack of or inadequate procedures (21 CFR 820.100(a))
Feb 20, 2020No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Aug 9, 2024Class IIZ-1292-2025Vios Monitoring System Bedside Monitor Model BSM2050
Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead…
Apr 22, 2024Class IIZ-1069-2025muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including…
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted…
Jun 6, 2023Class IIZ-1169-2025Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20…
A premature failure condition was identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor…
Jun 21, 2021Class IIZ-0956-2025muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including…
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted…

FDA record score

14 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 24, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

5 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIDec 11, 20246 x 16
Class II recall Z-1292-2025, this FEI as the recalling firmAug 9, 20244 x 0.62.4
Class II recall Z-1069-2025, this FEI as the recalling firmApr 22, 20244 x 0.62.4
Class II recall Z-1169-2025, this FEI as the recalling firmJun 6, 20234 x 0.62.4
Class II recall Z-0956-2025, this FEI as the recalling firmJun 21, 20214 x 0.251

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Murata Vios, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Murata Vios, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Murata Vios, Inc.
Location
Woodbury, MN (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.