Stryker Instruments FDA Inspection Record

FEI 3015967359 · Inspections, Form 483s, warning letters and import refusals

Stryker Instruments is an FDA-registered facility (FEI 3015967359) located in Portage, MI, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jan 24, 2023: Voluntary Action Indicated (VAI)
Inspections on record
4 (1 OAI, 1 VAI, 2 NAI)
Last OAI inspection
Sep 27, 2016
Warning letters
1 listed by FDA for this FEI, latest Apr 6, 2017
Recalls as the recalling firm
18, latest May 6, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jan 24, 2023Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Sep 27, 2016Official Action Indicated (OAI)No approval for changes requiring prior approval (21 CFR 812.35(a)(1))
Sponsor securing investigator compliance (21 CFR 812.46(a))
May 28, 2015No Action Indicated (NAI)No citations listed
Aug 20, 2012No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 18 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 6, 2026Class IIZ-2282-2026Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune…
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have…
Oct 31, 2025Class IIZ-0599-2026NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP; NN-8001 15x… and 2 more products
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause…
Jul 2, 2025Class IIZ-2277-2025Stryker SmartPump Tourniquet, disposable sterile (single use) DISP…
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting…
Feb 25, 2025Class IIZ-1469-2025Otopore Cylinder Standard outer ear dressings, Catalog Numbers 5400-010-000 and… and 13 more products
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile…
Jan 8, 2025Class IIZ-1138-2025stryker Precision Thin, Catalog Number REF 2296-003-108; Micro Oscillating and Sagittal… and 1 more product
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the…
Dec 3, 2024Class IIZ-0807-2025100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V…
Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers…
Oct 4, 2024Class IIZ-0365-2025Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for… and 5 more products
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are…
Sep 5, 2024Class IIZ-0063-20251.2mm Cross Cut Fissure - Intended for use in the cutting, drilling…
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only…
Jul 24, 2024Class IIZ-2722-2024Stryker iBur 3.0mm Precision Round, Distal Bend- Intended to cut bone in the following… and 9 more products
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures…
Jun 3, 2024Class IIZ-2259-2024The Non-Sterile Flyte hood covers the user s head and neck region and is intended to be… and 2 more products
Expired Products distributed to customers

FDA record score

64 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

21 records behind this number
RecordDatePoints x age weightAdds
Warning letter on FDA's dashboard, which does not show close-outsApr 6, 201730 x 0.257.5
Class II recall Z-2282-2026, this FEI as the recalling firmMay 6, 20264 x 14
Class II recall Z-0599-2026, this FEI as the recalling firmOct 31, 20254 x 14
Class II recall Z-2277-2025, this FEI as the recalling firmJul 2, 20254 x 14
Class II recall Z-1469-2025, this FEI as the recalling firmFeb 25, 20254 x 14
Class II recall Z-1138-2025, this FEI as the recalling firmJan 8, 20254 x 14
Class II recall Z-0807-2025, this FEI as the recalling firmDec 3, 20244 x 14
Inspection classified VAIJan 24, 20236 x 0.63.6
Class II recall Z-0365-2025, this FEI as the recalling firmOct 4, 20244 x 0.62.4
Class II recall Z-0063-2025, this FEI as the recalling firmSep 5, 20244 x 0.62.4
Class II recall Z-2722-2024, this FEI as the recalling firmJul 24, 20244 x 0.62.4
Class II recall Z-2259-2024, this FEI as the recalling firmJun 3, 20244 x 0.62.4
Class II recall Z-0239-2024, this FEI as the recalling firmSep 11, 20234 x 0.62.4
Class II recall Z-2596-2023, this FEI as the recalling firmAug 7, 20234 x 0.62.4
Class II recall Z-2463-2023, this FEI as the recalling firmJul 10, 20234 x 0.62.4
Class II recall Z-0980-2025, this FEI as the recalling firmJun 28, 20234 x 0.62.4
Class II recall Z-0981-2025, this FEI as the recalling firmJun 28, 20234 x 0.62.4
Class II recall Z-1339-2023, this FEI as the recalling firmMar 1, 20234 x 0.62.4
Class II recall Z-1184-2023, this FEI as the recalling firmJan 25, 20234 x 0.62.4
Class II recall Z-1478-2022, this FEI as the recalling firmJun 24, 20224 x 0.62.4
Inspection classified OAISep 27, 201610 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Stryker Instruments as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Stryker Instruments yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Stryker Corporation
Location
Portage, MI (US)
Source
device