Blue Belt Technologies, Inc. FDA Inspection Record
Blue Belt Technologies, Inc. is an FDA-registered facility (FEI 3016090213) located in Pittsburgh, PA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Feb 6, 2020: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (1 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest May 7, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Feb 6, 2020 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Evaluation, timeliness, identification (21 CFR 820.198(d)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Lack of Written MDR Procedures (21 CFR 803.17) and 2 more |
| Feb 24, 2016 | No Action Indicated (NAI) | No citations listed |
| Feb 6, 2013 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 7, 2024 | Class II | Z-2135-2024 | CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT - UNIVERSAL.Intended to assist… CORI Surgical System with Real Intelligence Software exhibits an increased Marker Registration Error (MRE) which may cause tracker arrays to… |
| Nov 13, 2023 | Class II | Z-0708-2024 | CORI REAL INTELLIGENCE Robotic Drill, Product No. ROB10013 and ROB10013S Possibility that the user is unable to resolve drill disconnection error messages. |
| Sep 7, 2023 | Class II | Z-0041-2024 | Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0… CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus… |
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2135-2024, this FEI as the recalling firm | May 7, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0708-2024, this FEI as the recalling firm | Nov 13, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0041-2024, this FEI as the recalling firm | Sep 7, 2023 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Feb 6, 2020 | 6 x 0.25 | 1.5 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 18, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 18, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 18, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing Blue Belt Technologies, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Blue Belt Technologies, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Blue Belt Technologies, Inc
- Location
- Pittsburgh, PA (US)
- Source
- device