PARKS MEDICAL ELECTRONICS, INC. FDA Inspection Record
PARKS MEDICAL ELECTRONICS, INC. is an FDA-registered facility (FEI 3020115) located in ALOHA, OR, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jun 15, 2023: Voluntary Action Indicated (VAI)
- Inspections on record
- 6 (1 OAI, 3 VAI, 2 NAI)
- Last OAI inspection
- Aug 17, 2012
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Nov 27, 2012. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Aug 8, 2012
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 15, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Jun 17, 2016 | No Action Indicated (NAI) | Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) Design verification - documentation (21 CFR 820.30(f)) Design validation - documentation (21 CFR 820.30(g)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| May 30, 2013 | Voluntary Action Indicated (VAI) | Software validation for automated processes (21 CFR 820.70(i)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) |
| Aug 17, 2012 | Official Action Indicated (OAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) DHR content (21 CFR 820.184) Servicing - Lack of or inadequate procedures (21 CFR 820.200(a)) and 6 more |
| Apr 20, 2012 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation - software validation documentation (21 CFR 820.30(g)) Design validation - Risk analysis (21 CFR 820.30(g)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 4 more |
| Jun 21, 2010 | No Action Indicated (NAI) | Quality Audit/Reaudit - conducted (21 CFR 820.22) Management review - defined interval, sufficient frequency (21 CFR 820.20(c)) Documentation (21 CFR 820.100(b)) |
Warning letters
- Nov 27, 2012: Warning letter listed on FDA's Data Dashboard. View compliance record (359639).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 8, 2012 | Class II | Z-0443-2013 | Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the… Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently… |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jun 15, 2023 | 6 x 0.6 | 3.6 |
| Inspection classified VAI | May 30, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Nov 27, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 17, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0443-2013, this FEI as the recalling firm | Aug 8, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 20, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing PARKS MEDICAL ELECTRONICS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check PARKS MEDICAL ELECTRONICS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Parks Medical Electronics, Inc
- Location
- ALOHA, OR (US)
- Source