WAVELY DIAGNOSTICS, INC. FDA Inspection Record
WAVELY DIAGNOSTICS, INC. is an FDA-registered facility (FEI 3022527636) located in Seattle, WA, US. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Apr 14, 2026: No Action Indicated (NAI)
- Inspections on record
- 1 (1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 14, 2026 | No Action Indicated (NAI) | External party controls do not include written quality agreements (ISO 13485:2016 Clause 4.1.5) Design and development validation not performed as planned (ISO 13485:2016 Clause 7.3.7) Supplier monitoring and re-evaluation not planned (ISO 13485:2016 Clause 7.4.1) Internal audits not conducted to ensure QMS meets requirements and is effective (ISO 13485:2016 Clause 8.2.4) |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. No OAI or VAI inspection, warning letter, Class I or II recall, or counted import refusals for this FEI in the records we hold.
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing WAVELY DIAGNOSTICS, INC. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for WAVELY DIAGNOSTICS, INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about WAVELY DIAGNOSTICS, INC. โ Check this supplier
Facility details
- Company
- Wavely Diagnostics, Inc.
- Location
- Seattle, WA (US)
- Source
- device