FDA Inspection 722658 — KORR Medical Technologies, Inc.

KORR Medical Technologies, Inc. — FEI 3002892786

The FDA completed an inspection of KORR Medical Technologies, Inc. on May 3, 2011 in Salt Lake City, UT. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 9 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
May 3, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Salt Lake City, UT (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471521 CFR 820.30(d)Design output - Lack of or inadequate procedures
1472021 CFR 820.50(a)(3)Acceptable supplier records, inadequate records
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
310221 CFR 820.30(h)Design transfer - Lack of or inadequate procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
317121 CFR 820.198(b)Rationale documented for no investigation
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures