FDA Inspection 878261 — KORR Medical Technologies, Inc.
The FDA completed an inspection of KORR Medical Technologies, Inc. on May 10, 2014 in Salt Lake City, UT. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 17 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- May 10, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Salt Lake City, UT (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 14715 | 21 CFR 820.30(d) | Design output - Lack of or inadequate procedures |
| 2350 | 21 CFR 820.25(b) | Training - Lack of or inadequate procedures |
| 2371 | 21 CFR 820.30(a) | Design control - no procedures |
| 3102 | 21 CFR 820.30(h) | Design transfer - Lack of or inadequate procedures |
| 3120 | 21 CFR 820.80(a) | Lack of or inadequate procedures - Acceptance activities |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3160 | 21 CFR 820.184 | Lack of or inadequate DHR procedures |
| 3175 | 21 CFR 820.186 | QSR |
| 3207 | 21 CFR 820.50(b) | Supplier notification of changes |
| 3233 | 21 CFR 820.72(a) | Calibration, Inspection, etc. Procedures Lack of or Inadequ |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 3286 | 21 CFR 820.90(b)(1) | Procedures for product review,disposition lack of/inadequate |
| 3666 | 21 CFR 820.20(c) | Management review - Lack of or inadequate procedures |
| 447 | 21 CFR 820.40 | Lack of procedures, or not maintained |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |