FDA Inspection 807507 — Hill-Rom Inc

Hill-Rom Inc — FEI 1824206

The FDA completed an inspection of Hill-Rom Inc on Jan 11, 2013 in Batesville, IN. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 6 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Official Action Indicated (OAI)
End Date
Jan 11, 2013
Fiscal Year
2013
Product Type
Devices
Project Area
Compliance: Devices
Location
Batesville, IN (United States)

Citations

IDCFRDescription
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
265021 CFR 820.30(f)Design verification - Lack of or inadequate procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
317221 CFR 820.198(c)Investigation of device failures
50221 CFR 820.60Identification procedures, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures