FDA Inspection 865826 — Hill-Rom Inc
The FDA completed an inspection of Hill-Rom Inc on Feb 13, 2014 in Batesville, IN. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 8 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Official Action Indicated (OAI)
- End Date
- Feb 13, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Batesville, IN (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 14716 | 21 CFR 820.30(f) | Design verification - output does not meet input requirement |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3172 | 21 CFR 820.198(c) | Investigation of device failures |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 3285 | 21 CFR 820.90(b)(2) | Product rework procedures, Lack of or inadequate procedures |
| 3680 | 21 CFR 820.70(a) | Process control procedures, Lack of or inadequate procedures |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |