HILL-ROM, INC. FDA Inspection Record
HILL-ROM, INC. is an FDA-registered facility (FEI 1824206) located in Batesville, IN, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 12, 2024: No Action Indicated (NAI)
- Inspections on record
- 8 (3 OAI, 4 VAI, 1 NAI)
- Last OAI inspection
- Feb 13, 2014
- Warning letters
- 1 listed by FDA for this FEI, latest Mar 2, 2012
- Form 483 and inspection documents
- 1 in FDA's reading room
- Recalls as the recalling firm
- 26, latest Jan 14, 2022
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 12, 2024 | No Action Indicated (NAI) | No citations listed |
| Nov 19, 2021 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) |
| Dec 19, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of Written MDR Procedures (21 CFR 803.17) |
| Feb 13, 2014 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) and 4 more |
| Jan 11, 2013 | Official Action Indicated (OAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Investigation of device failures (21 CFR 820.198(c)) and 2 more |
| Oct 11, 2011 | Official Action Indicated (OAI) | No citations listed |
| Jul 6, 2010 | Voluntary Action Indicated (VAI) | Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Acceptance status (21 CFR 820.86) |
| Jan 23, 2009 | Voluntary Action Indicated (VAI) | Identifying corrective & preventive actions (21 CFR 820.100(a)(3)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 14, 2022 | Class II | Z-0783-2022 | (1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead… Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the… |
| Dec 23, 2021 | Class II | Z-0639-2022 | Liko Mobile Lifts: Primarily intended for use in nursing homes
Liko M220, Model 2050010 and 1 more product Missing bushing on Liko M220 and M230 causing wear and potential for patient falling |
| Aug 11, 2021 | Class II | Z-2597-2021 | OctoStretch with Stretch Leveller - The combination of this products allows the user to… and 4 more products Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion… |
| May 28, 2021 | Class II | Z-2156-2021 | Affinity Four Birthing Bed-used as a birthing bed for women of child bearing age in an LDR… Latch mechanism on the Lift-Off foot section of the Affinity Four Birthing Bed is damaged, it could potentially cause the installed foot… |
| Dec 18, 2020 | Class I | Z-0925-2021 | Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the… Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming… |
| Jun 30, 2020 | Class II | Z-2737-2020 | Progressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent… and 1 more product Affected beds may have screws installed that do not meet material tensile strength requirements. Failure of this screw could result in… |
| Apr 30, 2020 | Class II | Z-2246-2020 | Centrella Smart+ Bed, Catalog Number P7900 The Bed Exit System may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the… |
| Jan 12, 2018 | Class II | Z-1319-2018 | Progressa Bed - AC-powered adjustable hospital bed A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged… |
| Feb 23, 2017 | Class II | Z-1816-2017 | TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient… The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm… |
| Aug 9, 2016 | Class II | Z-0326-2017 | Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling
installed rail… Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a… |
Form 483s and inspection documents
FDA record score
16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
34 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0925-2021, this FEI as the recalling firm | Dec 18, 2020 | 15 x 0.25 | 3.8 |
| Class II recall Z-0783-2022, this FEI as the recalling firm | Jan 14, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0639-2022, this FEI as the recalling firm | Dec 23, 2021 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Nov 19, 2021 | 6 x 0.25 | 1.5 |
| Class II recall Z-2597-2021, this FEI as the recalling firm | Aug 11, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2156-2021, this FEI as the recalling firm | May 28, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2737-2020, this FEI as the recalling firm | Jun 30, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2246-2020, this FEI as the recalling firm | Apr 30, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1319-2018, this FEI as the recalling firm | Jan 12, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1816-2017, this FEI as the recalling firm | Feb 23, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0326-2017, this FEI as the recalling firm | Aug 9, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2750-2016, this FEI as the recalling firm | Jul 15, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2376-2016, this FEI as the recalling firm | Jun 14, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1916-2016, this FEI as the recalling firm | Mar 8, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0860-2016, this FEI as the recalling firm | Jan 20, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0274-2016, this FEI as the recalling firm | Oct 7, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1818-2015, this FEI as the recalling firm | Mar 20, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1820-2015, this FEI as the recalling firm | Mar 2, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 19, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2562-2014, this FEI as the recalling firm | Jun 26, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Feb 13, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1794-2013, this FEI as the recalling firm | May 30, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1473-2013, this FEI as the recalling firm | Apr 30, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1466-2013, this FEI as the recalling firm | Apr 15, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0949-2013, this FEI as the recalling firm | Feb 26, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jan 11, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0622-2013, this FEI as the recalling firm | Dec 6, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0872-2013, this FEI as the recalling firm | Oct 31, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1864-2012, this FEI as the recalling firm | Jun 18, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2082-2012, this FEI as the recalling firm | Apr 2, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Mar 2, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 11, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 6, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 23, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |