Z-0001-2014 Class II Terminated

Recalled by Maquet Cardiovascular, LLC — Wayne, NJ

FDA device recall Z-0001-2014 was initiated by Maquet Cardiovascular, LLC on Sep 20, 2013 and is designated Class II. Reason for recall: MAQUET has identified a potential issue with the ULTIMA OPCAB (part numbers OM-2003S & OM-2001D). Labeling for the 2 blade lots had been incorrectly applied. The standard blades from Lot #25052176… The recall status is terminated (terminated Jun 10, 2014). Affected quantity: 20 units (Lot 25052176); 12 units (Lot #25052200).

Recall details

Product Type
Devices
Report Date
Oct 9, 2013
Initiation Date
Sep 20, 2013
Termination Date
Jun 10, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 units (Lot 25052176); 12 units (Lot #25052200)

Product description

Ultima OPCAB System, Sterile, Rx Only, Product Usage: Intended for use during performance of minimally invasive cardiac surgery through a sternotomy incision approach. The ULTIMA Access RailPlatform in combination with the Activator II Drive Mechanism is used to spread the sternum, providing access and direct visualization to the thoracic cavity. The AccessRail Platform also allows for the organization of pericardial sutures. The ULTIMA Stablizer isolates and provides local immobilization of the target vessel on the beating heart.

Reason for recall

MAQUET has identified a potential issue with the ULTIMA OPCAB (part numbers OM-2003S & OM-2001D). Labeling for the 2 blade lots had been incorrectly applied. The standard blades from Lot #25052176 were incorrectly labeled as deep blades, and the 20 deep blades from Lot #25052200 were incorrectly labeled as standard blades.

Distribution pattern

USA Nationwide Distribution in the states of: Florida, Maryland, New Jersey, New York and Texas.

Code information

ULTIMA OPCAB System Part Nos. OM-2003S (LOT #25052176) and OM-2001D (LOT #25052200).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0001-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0001-2014. Report a data issue: hello@regdatalab.com.