Z-0004-2025 Class I Ongoing

Recalled by Mercury Enterprises, Inc. Dba Mercury Medical — Clearwater, FL

FDA device recall Z-0004-2025 was initiated by Mercury Enterprises, Inc. Dba Mercury Medical on Aug 30, 2024 and is designated Class I. Reason for recall: Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step. The recall status is ongoing. Affected quantity: 1,300 units.

Recall details

Product Type
Devices
Report Date
Oct 9, 2024
Initiation Date
Aug 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,300 units

Product description

Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.

Reason for recall

Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step.

Distribution pattern

US distribution to MI, FL, SC, PA, IN. International distribution to Albania.

Code information

Part Number: 1050805; UDI/DI: 10641043508053, 30641043508057; Lot Numbers: 2417650805, 2417750805. Part Number: 1050832; UDI/DI: 10641043508329, 30641043508323; Lot Numbers: 2423550832, 2417850832.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0004-2025
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Mercury Enterprises, Inc. Dba Mercury Medical? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0004-2025. Report a data issue: hello@regdatalab.com.