Z-0008-2023 Class II Ongoing

Recalled by Synthes (USA) Products LLC — West Chester, PA

FDA device recall Z-0008-2023 was initiated by Synthes (USA) Products LLC on Aug 30, 2022 and is designated Class II. Reason for recall: Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is… The recall status is ongoing. Affected quantity: 191 units.

Recall Details

Product Type
Devices
Report Date
Oct 12, 2022
Initiation Date
Aug 30, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
191 units

Product Description

Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001

Reason for Recall

Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: 00810633022047 Lot Numbers: MHC220119, MHC220120 Expiry Date: 1 June 2027

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0008-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.