Z-1916-2024 Class II Ongoing

Recalled by Synthes (USA) Products LLC — West Chester, PA

FDA device recall Z-1916-2024 was initiated by Synthes (USA) Products LLC on Apr 25, 2024 and is designated Class II. Reason for recall: Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay The recall status is ongoing. Affected quantity: 261 units.

Recall details

Product Type
Devices
Report Date
Jun 5, 2024
Initiation Date
Apr 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
261 units

Product description

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

Reason for recall

Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

Distribution pattern

Worldwide distribution - US Nationwide and the country of Canada.

Code information

UDI-DI: 10886982075192 Lot Numbers: 97P6250, 97P6251, 97P6252

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1916-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1916-2024. Report a data issue: hello@regdatalab.com.