Z-1702-2026 Class II Ongoing
FDA device recall Z-1702-2026 was initiated by Synthes (USA) Products LLC on February 25, 2026 and is designated Class II. Reason for recall: A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread. The recall status is ongoing. Affected quantity: 77 units.
Recall Details
- Product Type
- Devices
- Report Date
- April 8, 2026
- Initiation Date
- February 25, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 77 units
Product Description
2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.
Reason for Recall
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.
Distribution Pattern
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.
Code Information
Part Number: 03.424.124-US. GTIN: 10886982335944. Lot Number: 68900P9.