Z-0010-2019 Class II Terminated

Recalled by Radiometer America Inc — Brea, CA

Recall Details

Product Type
Devices
Report Date
October 10, 2018
Initiation Date
May 25, 2018
Termination Date
November 6, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12,102

Product Description

ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Reason for Recall

The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse ¿ Immediately after lifts the inlet and quickly closes it again

Distribution Pattern

US Distribution and Internationally to Canada.

Code Information

ABL90 FLEX analyzers with software versions 3.3MR1 or below.