Z-0010-2019 Class II Terminated
FDA device recall Z-0010-2019 was initiated by Radiometer America Inc on May 25, 2018 and is designated Class II. Reason for recall: The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby… The recall status is terminated (terminated Nov 6, 2020). Affected quantity: 12,102.