Z-0010-2019 Class II Terminated

Recalled by Radiometer America Inc — Brea, CA

FDA device recall Z-0010-2019 was initiated by Radiometer America Inc on May 25, 2018 and is designated Class II. Reason for recall: The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby… The recall status is terminated (terminated Nov 6, 2020). Affected quantity: 12,102.

Recall details

Product Type
Devices
Report Date
Oct 10, 2018
Initiation Date
May 25, 2018
Termination Date
Nov 6, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12,102

Product description

ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Reason for recall

The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse ¿ Immediately after lifts the inlet and quickly closes it again

Distribution pattern

US Distribution and Internationally to Canada.

Code information

ABL90 FLEX analyzers with software versions 3.3MR1 or below.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0010-2019
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Radiometer America Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0010-2019. Report a data issue: hello@regdatalab.com.