Z-0016-2018 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-0016-2018 was initiated by Philips Electronics North America Corporation on Jan 18, 2017 and is designated Class II. Reason for recall: During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the… The recall status is terminated (terminated Feb 26, 2020). Affected quantity: US - 606.

Recall details

Product Type
Devices
Report Date
Oct 25, 2017
Initiation Date
Jan 18, 2017
Termination Date
Feb 26, 2020
Voluntary/Mandated
FDA Mandated
Product Quantity
US - 606

Product description

Philips Healthcare Ingenuity Core X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason for recall

During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.

Distribution pattern

USA (nationwide) Distribution

Code information

Software version 4.16

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0016-2018
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0016-2018. Report a data issue: hello@regdatalab.com.