Z-0020-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 25, 2017
- Initiation Date
- January 18, 2017
- Termination Date
- February 26, 2020
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- US - 606
Product Description
Philips Healthcare Brilliance iCT SP X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason for Recall
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
Distribution Pattern
USA (nationwide) Distribution
Code Information
Software 4.16