Z-0025-2016 Class II Terminated
FDA device recall Z-0025-2016 was initiated by Integra LifeSciences Corp. on Aug 21, 2015 and is designated Class II. Reason for recall: Integra identified that the UCHRAP component (Arc Adapter Plate) would not assemble properly to the UCHRAR component (ARC Adapter Ring) of the Compact Head Ring Adapter due to an error in the… The recall status is terminated (terminated Oct 13, 2016). Affected quantity: 6 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 14, 2015
- Initiation Date
- Aug 21, 2015
- Termination Date
- Oct 13, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6 units
Product description
Universal Compact Head Ring Adapter Plate (UCHRAP), a component of the Universal Compact Head Ring Adapter, UCHRA. The UCHRA is a component of the Integra¿ CRW¿ Precision Arc Stereotactic System.
Reason for recall
Integra identified that the UCHRAP component (Arc Adapter Plate) would not assemble properly to the UCHRAR component (ARC Adapter Ring) of the Compact Head Ring Adapter due to an error in the manufacturing drawing.
Distribution pattern
Worldwide Distribution - US Distribution to the states of : NJ, TX and WY., and to the countries of : Algeria, Germany and India.
Code information
Cat. No. UCHRAP; Serial No.s 900, 901, 902, 904, 905, 906