Z-0028-2022 Class II Terminated

Recalled by Johnson & Johnson Surgical Vision Inc — Santa Ana, CA

FDA device recall Z-0028-2022 was initiated by Johnson & Johnson Surgical Vision Inc on Aug 6, 2021 and is designated Class II. Reason for recall: Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin… The recall status is terminated (terminated Mar 6, 2023). Affected quantity: 1 lens.

Recall details

Product Type
Devices
Report Date
Oct 13, 2021
Initiation Date
Aug 6, 2021
Termination Date
Mar 6, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 lens

Product description

TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. The TECNIS Simplicity Delivery System is designed as a single-use disposable preloaded modular cartridge that functions as both the primary packaging and as a sterile, disposable insertion system for delivering the IOL into the eye during cataract surgery.

Reason for recall

Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit. Product was released in error based on the results of additional testing performed during the investigation into release testing nonconformance.

Distribution pattern

U.S. Nationwide distribution in the state of IL.

Code information

Model: DIU225U050 UDI Code: (01)05050474745831(17)240414(21)5248142104 Production Order Number: 900001985625 Serial Number: 5248142104

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0028-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Johnson & Johnson Surgical Vision Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Johnson & Johnson Surgical Vision Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Johnson & Johnson Surgical Vision Inc →

Source: openFDA record for Z-0028-2022. Report a data issue: hello@regdatalab.com.