Z-0028-2026 Class I Ongoing

Recalled by NOXBOX LTD — Sittingbourne

Recall Details

Product Type
Devices
Report Date
October 15, 2025
Initiation Date
September 9, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1667 units

Product Description

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Reason for Recall

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Distribution Pattern

Domestic: TN

Code Information

Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.