Z-0028-2026 Class I Ongoing
FDA device recall Z-0028-2026 was initiated by NOXBOX LTD on Sep 9, 2025 and is designated Class I. Reason for recall: Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose… The recall status is ongoing. Affected quantity: 1667 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 15, 2025
- Initiation Date
- Sep 9, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1667 units
Product description
Reason for recall
Distribution pattern
Domestic: TN
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0028-2026
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026