Z-0035-2026 Class I Ongoing

Recalled by CareFusion 303, Inc. — San Diego, CA

Recall Details

Product Type
Devices
Report Date
October 15, 2025
Initiation Date
September 11, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity

Product Description

BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite Y-Site 10015414, 180 Micron Filter 2447-0007, Ball Valve SmartSite Port (Burette) 2 SmartSites Y-Sites 24201-0007, 2 SmartSite Y-sites 10012645, Back Check Valve, Non-vented, SmartSite Y-site 2426-0500, Back Check Valve, 3 SmartSite Y-sites 2450-0500, Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSite Y-sites 11522558, Ball Valve, Back Check Valve, 3 SmartSite Y-sites 2410-0500, 2 SmartSite Y-Sites 2232-0007, 0.2 Micron Filter 11613191, 0.2 Micron Filter, Ball Valve, SmartSite" Port (Burette), 2 SmartSite Y-sites 10012144, 3 SmartSite Y-Sites, Check Valve 10013034, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 4-way Stopcock 10013037, Non-Vented, 180 Micron Filter, Low Sorbing Tubing Segment, 1 SmartSite, Y-Site 11582773, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 3-way Stopcock

Reason for Recall

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

Distribution Pattern

US: Nationwide OUS: United Arab Emirates Australia Belgium Bahrain Botswana Canada Switzerland China France United Kingdom Gibraltar Israel India Kuwait Mexico Malaysia Netherlands New Zealand Philippines Pakistan Qatar Saudi Arabia Singapore Taiwan, Province of China South Africa

Code Information

All Lots/ UDI: C24101E - 10885403235528, 10015414 - 10885403233999, 2447-0007 - 10885403235078, 24201-0007 - 10885403476341, 10012645 - 10885403233753, 2426-0500 - 7613203020992, 11522558 - 10885403232305, 2410-0500 - 7613203012492, 2232-0007 - 10885403463969, 11613191 - 10885403276026, 10012144 - 7613203021159, 10013034 - 10885403232312, 10013037 - 10885403233784, 11582773 - 10885403239588