Z-0037-2024 Class II Ongoing

Recalled by AB Sciex, LLC — Framingham, MA

Recall Details

Product Type
Devices
Report Date
October 11, 2023
Initiation Date
August 30, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

API3200MD Mass Spectrometer, Part Number 4466230

Reason for Recall

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Distribution Pattern

US Nationwide.

Code Information

Serial Numbers: EQ230011907, EQ230021911, EQ230001901