Z-0041-2024 Class II Ongoing

Recalled by Blue Belt Technologies, Inc — Pittsburgh, PA

FDA device recall Z-0041-2024 was initiated by Blue Belt Technologies, Inc on Sep 7, 2023 and is designated Class II. Reason for recall: CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the… The recall status is ongoing. Affected quantity: 56 units.

Recall details

Product Type
Devices
Report Date
Oct 11, 2023
Initiation Date
Sep 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
56 units

Product description

Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

Reason for recall

CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis

Distribution pattern

US Nationwide distribution.

Code information

UDI: (01) 00885556773338 Lot Numbers: 51141585, 51108475, 51108689

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0041-2024
openFDA file date
Nov 17, 2026
Loaded by RegDataLab
Nov 18, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0041-2024. Report a data issue: hello@regdatalab.com.