Z-0041-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 11, 2023
- Initiation Date
- September 7, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 56 units
Product Description
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
Reason for Recall
CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis
Distribution Pattern
US Nationwide distribution.
Code Information
UDI: (01) 00885556773338 Lot Numbers: 51141585, 51108475, 51108689