Z-0043-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 16, 2024
- Initiation Date
- September 5, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 613
Product Description
NxTAG Respiratory Pathogen Panel +SARS-CoV-2, REF: I056C0468
Reason for Recall
Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.
Distribution Pattern
U Nationwide distribution in the states of OH, CA, NC, WI, TN, TX, MI, NY, MO, FL, MA, NJ, PA, VA, AZ, IL, OK, CO, MN.
Code Information
Lot/Expiration: IK056C-0033/ 9/30/2024, IK056C-0034/ 10/31/2024, IK056C-0035/ 10/31/2024, IK056C-0036/ 12/31/2024, IK056C-0037/ 12/31/2024, IK056C-0038/ 12/31/2024, IK056C-0039/ 1/31/2025