Z-0044-2014 Class II Terminated

Recalled by Advanced Sterilization Products — Irvine, CA

FDA device recall Z-0044-2014 was initiated by Advanced Sterilization Products on Sep 18, 2013 and is designated Class II. Reason for recall: Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for… The recall status is terminated (terminated Jan 13, 2014). Affected quantity: 2,995 units.

Recall details

Product Type
Devices
Report Date
Oct 30, 2013
Initiation Date
Sep 18, 2013
Termination Date
Jan 13, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,995 units

Product description

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Reason for recall

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

Distribution pattern

Nationwide Distribution

Code information

Lot# 13F050, 13F051, 13F049

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0044-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Advanced Sterilization Products? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Advanced Sterilization Products: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Advanced Sterilization Products →

Source: openFDA record for Z-0044-2014. Report a data issue: hello@regdatalab.com.