Z-0047-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc. — Hoffman Estates, IL

Recall Details

Product Type
Devices
Report Date
November 1, 2017
Initiation Date
September 18, 2017
Termination Date
December 11, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 devices

Product Description

Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.

Reason for Recall

The torque wrench used to tighten system bolts during installation was found to be out of tolerance

Distribution Pattern

Distribution was made to medical facilities located in CA, FL, GA, LA, NY, PA, and TN. There was government distribution but no foreign/military distribution.

Code Information

Symbia T2 serial number 2001; Symbia Intevo Excel serial numbers 2112 and 2140; Symbia Intevo 16 serial numbers 2100 and 2116; Symbia Evo serial numbers 1049, 1054, 1071, 1075, 1104, and 1128; and Symbia Evo Excel serial numbers 1109, 1111, 1131, 1134, and 1158.