Z-0048-2025 Class I Ongoing

Recalled by DATEX--OHMEDA, INC. — Wauwatosa, WI

FDA device recall Z-0048-2025 was initiated by DATEX--OHMEDA, INC. on Sep 3, 2024 and is designated Class I. Reason for recall: GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Oct 23, 2024
Initiation Date
Sep 3, 2024
Voluntary/Mandated
Voluntary: Firm initiated

Product description

GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator

Reason for recall

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

Distribution pattern

Worldwide distribution.

Code information

All models

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0048-2025
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for DATEX--OHMEDA, INC.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for DATEX--OHMEDA, INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about DATEX--OHMEDA, INC. →

Source: openFDA record for Z-0048-2025. Report a data issue: hello@regdatalab.com.