Z-0050-2024 Class II Ongoing
Recalled by Universal Meditech Inc.
— Reedley, CA
FDA device recall Z-0050-2024 was initiated by Universal Meditech Inc. on May 22, 2023 and is designated Class II. Reason for recall: Universal Meditech Inc. was violatively distributing PrestiBio" Breastmilk Alcohol Test Strip, DiagnosUS¿ SARS-CoV-2 Antibody (IgG/IgM) Test, HealthyWiser KetoFast" Ketone Test Strips and PrestiBio"… The recall status is ongoing. Affected quantity: 48 boxes (50 strips/box).
Recall details
- Product Type
- Devices
- Report Date
- Nov 1, 2023
- Initiation Date
- May 22, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 48 boxes (50 strips/box)
Product description
To Life hCG Pregnancy Urine Test Strips Format REF Cat No: 100-10
Reason for recall
Universal Meditech Inc. was violatively distributing PrestiBio" Breastmilk Alcohol Test Strip, DiagnosUS¿ SARS-CoV-2 Antibody (IgG/IgM) Test, HealthyWiser KetoFast" Ketone Test Strips and PrestiBio" Ketone Test Strips without marketing authorization. UMI is also recall their tests because they are going out of business and would not be able to continue fulfilling any post-market responsibilities of these distributed products. UMI claims to hold 510(k)s for the other devices, purchased the intellectual property in 2015 from the previous 510(k) holder, but FDA has been unable to verify this information. The devices may have unknown performance characteristics because the firm is unable to provide any documentation that the devices were manufactured in conformance with the Quality System regulation, including documentation (e.g., DHF) that the distributed devices had not been modified since original clearance in a way that could impact their safety and effectiveness, documentation of controlled storage temperature/humidity, and post-market surveillance documentation, such as complaints and adverse events.
Distribution pattern
U.S.: CA, FL, IL, LA, NJ, NY, TX and WY O.U.S.: Taiwan
Code information
Model Number: 100-13 UDI-DI Code: None Lot Numbers: hCG-S-2210-03
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0050-2024
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Universal Meditech Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Universal Meditech Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Universal Meditech Inc. →
Use it in ChatGPT
Get a free API key
How the connector works