Z-0051-2013 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-0051-2013 was initiated by Terumo Cardiovascular Systems Corporation on Sep 24, 2012 and is designated Class II. Reason for recall: Terumo Cardiovascular Systems (Terumo CVS) received one report of loss of power to the control module and subsequently the drive motor for the Sarns centrifugal system during cardiopulmonary bypass. The recall status is terminated (terminated Nov 13, 2013). Affected quantity: 847 units.

Recall details

Product Type
Devices
Report Date
Oct 17, 2012
Initiation Date
Sep 24, 2012
Termination Date
Nov 13, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
847 units

Product description

Sarns Centrifugal System Control module, 100/115V The Sarns Centrifugal System is indicated for use in cardiopulmonary bypass procedures only.

Reason for recall

Terumo Cardiovascular Systems (Terumo CVS) received one report of loss of power to the control module and subsequently the drive motor for the Sarns centrifugal system during cardiopulmonary bypass.

Distribution pattern

Worldwide Distribution - USA (nationwide) and the countries of Bangladesh, Seoul Korea, Chile, Brazil, Venesuela, Thailand, Mexico, India, Taiwan, Canada, Dominican Republic, Guatemala, Ecuador, Germany, Republic of Georgia, Hong Kong, Japan, United Arab Emirates (UAE), Japan, Belgium, Phillipines, Singapore, and South Africa.

Code information

Catalog number: 6379 and serial numbers: 942,1003-1006,1034,1039,1051,1056,1087,1914,3030,4019,4344, 5009,5012,5013,5030,5048,5059, 5062, 5080, 5096, 5111, 5112, 5122, 5140, 5713, 6003, 6005, 6023, 6024, 6027, 6033, 6034, 6054, 6055, 6060, 6066, 6068, 6069, 6073, 6079, 6081, 6082, 6086, 6089, 6090-6092, 6097, 6107, 6109, 6112, 6119, 6126-6128, 6236, 6379, 7000-7785, and R7051.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0051-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-0051-2013. Report a data issue: hello@regdatalab.com.