Z-0051-2025 Class II Ongoing

Recalled by AURIS HEALTH INC — Redwood City, CA

Recall Details

Product Type
Devices
Report Date
October 16, 2024
Initiation Date
August 2, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
176

Product Description

Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, MON-000006-RFB.

Reason for Recall

Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.

Code Information

Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00. REF/UDI-DI/Serial Numbers: MON-000005-01/+B634MON000005010/110024, 13, 110014, 110016, 110040, 110038, 110027, 110035, 110022, 110034, 1806P2, 120170, 110045, 110039, 110049, 110047, 110052, 110030, 110032, 110021, 110043, 110050, 110029, 110033, 110046, 110048, 110018, 110015, 110051, 1806P1, 110012, 14, 110041, 110036; MON-000005-01R/+B634MON00000501R0/110020, 110017, 110031, 110028, 120042, 110013. MON-000006/+B634MON0000060/120088, 120057, 120046, 120058, 120121, 120152, 120015, 120160, 120068, 120167, 120025, 120030, 120044, 120093, 120040, 120063, 120092, 120173, 120036, 120064, 120122, 120056, 120128, 120017, 120089, 120075, 120031, 120169, 120113, 120168, 120055, 120148, 120014, 120060, 120124, 120164, 120123, 120007, 120006, 120004, 120120, 120126, 120082, 120047, 120051, 120020, 120103, 120153, 120108, 120037, 120074, 120005, 120130, 120095, 120096, 120009, 120111, 120022, 120054, 120163, 120027, 120090, 120141, 120072, 120155, 120101, 120143, 120125, 120052, 120100, 120050, 120098, 120045, 120131, 120013, 120105, 120165, 120136, 120029, 120041, 120139, 120137, 120053, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 120145, 120127, 120011, 120070, 120076, 120134, 120117, 120077, 120149, 120102, 120002, 120010, 120003, 120023, 120146, 120043, 120071, 120107, 120065, 120116, 120150, 120066, 120166, 120033, 120079, 120008, 120132, 120086, 120032, 120073, 120016, 120080, 120078, 120012, 120084, 120083, 120038, 120067, 120021; MON-000006-RFB/+B634MON0000060/120081, 120026, 120091, 120106, 120104.