Z-0052-2021 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

Recall Details

Product Type
Devices
Report Date
October 14, 2020
Initiation Date
August 19, 2020
Termination Date
August 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
533 US and 1822 OUS

Product Description

Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068

Reason for Recall

Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed

Distribution Pattern

US Nationwide distribution.

Code Information

Lot # EB0255 UDI: 00842768037726EB025520091110471068840