Z-0061-2019 Class II Ongoing

Recalled by Ortho Clinical Diagnostics Inc — Raritan, NJ

Recall Details

Product Type
Devices
Report Date
October 10, 2018
Initiation Date
July 12, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
68230

Product Description

VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997

Reason for Recall

The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.

Distribution Pattern

Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.

Code Information

5535 5540 5550 5560 5570 5580 5590 5600 5610 5615 5620 5630 5640 5650 5655 5660 5665 5670 5675 5678 5680 5685 5690