Z-0844-2021 Class II Terminated

Recalled by Ortho Clinical Diagnostics Inc — Raritan, NJ

Recall Details

Product Type
Devices
Report Date
January 20, 2021
Initiation Date
December 15, 2020
Termination Date
July 27, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US= 223,788 ; OUS:=168,120 Totlal: 491,908

Product Description

VITROS Chemistry Products TP Slides- IVD measure total protein (TP) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems Product Code: 8392292

Reason for Recall

Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of AR, AU, BM, BR, CA, CL, CN, CO, CR, DO, EC, FR, FR, GT, HT, IN, JP, MX, PA, PE, PR, PY, SV.

Code Information

All Current, Future and Past Lots UDI: 10758750004850