Z-0471-2021 Class III Ongoing

Recalled by Ortho Clinical Diagnostics Inc — Raritan, NJ

Recall Details

Product Type
Devices
Report Date
November 25, 2020
Initiation Date
October 16, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10788 total: US 4680; OUS 6108

Product Description

VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474

Reason for Recall

Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Code Information

Lot Q7692 , EXP: 01-NOV-2021 UDI: 10758750004577