Z-0063-2018 Class II Ongoing

Recalled by North Star Imaging Inc — ROGERS, MN

FDA device recall Z-0063-2018 was initiated by North Star Imaging Inc on August 23, 2017 and is designated Class II. Reason for recall: It was discovered a low-level radiation leak that was found on the NSI X-7000 system. The recall status is ongoing. Affected quantity: US - 3 units.

Recall Details

Product Type
Devices
Report Date
November 15, 2017
Initiation Date
August 23, 2017
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
US - 3 units

Product Description

North Star Imaging Inc. (NSI) Model X-7000 industrial cabinet x-ray system. Industrial x-ray imaging of product hardware

Reason for Recall

It was discovered a low-level radiation leak that was found on the NSI X-7000 system.

Distribution Pattern

US Distribution to the states of : PA, OH and KS.

Code Information

Model X-7000